Data Acquisition System Logger StationⅡ DL2800
32-Channels with Multiple Sensor Type Direct Input
Control and Measure Both Stand-alone & via Network
Features
Remote Control Measurement & Recording via Network
Up to eight (8) DL2800 units (256 channels) can be remotely Network controlled via Ethernet connection.
Multi-channel Measurements
Each mainframe provides 32 Channel capacity via 16 dual channel AP-Amplifier Module slots. Additionally, 16 Logic/Event Digital Inputs are provided standard.
Stand-alone Operation
Simply connect user's monitor, keyboard and USB mouse to operate the DL2800 as a selfcontained stand-alone Data Logger
Common Signal Amplifiers
Plug-in Signal Input Conditioning Amplifier Modules are the same as used with the RA1000 and RA2000 series Data Acquisition Recorders. Each isolated Amplifier can be interchanged among units for measurement versatility.
Three (3) Powerful Measurement Modes
- Memory Mode - Record Fast Events
- HDD Mode - Long-term Measurements
- Multi Mode - Combined Memory & HDD
Rack Mounting Available
Optional 19"rack mount adapters are available for the DL2800 mainframe. A front panel located HDD simplifies maintenance.
Direct Input from Various Sensors
Plug-in AP Series Amplifiers used with RA recorders enable direct sensor input for Voltage, Vibration, Strain, Temperature, Pressure, Pulse, etc.
Multi-unit Synchronization Option
Time synchronize up to 10 each DL2800 mainframes (320 channels) with this option when operating in the Memory Recorder mode.
Make Both Transient and Longterm Recordings
Transients: Up to 2µS sampling Long-term: Record for 60-days at 10mS sampling on 32 channels.
Basic Specifications
| Number of channels | 32 channels (16 slots) + 16 digital input (mainframe) | ||
|---|---|---|---|
| Storage device | Internal memory: 32MW,Built-in HDD: 40GB (includes system domain 5GB) USB memory using the USB port |
||
| Interfaces | Ethernet (10BASE-T/100BASE-TX): for control with communication command (std) USB mouse: for USB memory, 2 ports (standard) RS-232C: for control by communication command or UPS shut-down (optional) |
||
| Sampling speed | Max 2µs (500kS/s) **Depends on recording drive and number of channels | ||
| Measuring mode | Selectable from Memory Recorder, HDD Recorder and Multi Recorder modes | ||
| Memory Recorder Mode | Records measured data onto internal memory by set trigger. For fast or accidental events | ||
| Memory capacity | 32MW (1MW per channel when using 32 channels) | ||
| Memory division | 1, 2, 4, 8, 16, 32, 64, 128 divisions | ||
| Measuring Operation | One time, repeat, endless | ||
| HDD Recorder Mode | Records measured data on HDD at real time. For long-term measurement. | ||
| HDD capacity | Max 35GB | ||
| Starting condition | Press start key or set start time. | ||
| Measuring Operation | Normal or ring (repeat recording at set time period) | ||
| Recording method | Sample: record data at set sampling speed Peak: record max and min value at every 2µs sampling during set recording period |
||
| Multi Recorder Mode | Combination of Memory & HDD recording. Steady-state and transient events can be recorded simultaneously. | ||
| Monitor output | XGA (analog RGB, 1024 x 768 dots) | ||
| Operating environment | Temperature: 5 to 40℃, humidity: 35 to 80%RH (Without condensation. Humidity must be 45%RH or lower if temperature is 30℃ or over.) |
||
| Power supply | 90 to 264VAC, frequency: 50 or 60Hz | ||
| Power consumption | Max approx. 170VAC (when using USB port and 16 AP11-103s installed) | ||
| Dimensions | Approx. 431 (W) x 244 (H) x 307 (D) (pads on corners included but not other projections) | ||
| Weight | Mainframe only: 10.5kg or less Mainframe, 1 optional unit & 12 AP11-101s: 13.5kg or less |
||
Connection with PC
Data recorded on DL2800 can be transferred to the PC via USB memory or Ethernet. The optional PC software ‘Unifizer’allows users to do calculation, report generation and file conversion of the data.

Recordable Time on Memory Devices
| Sampling Speed | Internal memory (32MW) | HDD (35GB) *1 | USB memory (512MB) *1 | |||
|---|---|---|---|---|---|---|
| w/ 1ch | w/ 32chs | w/ 1ch | w/ 32chs | w/ 1ch | w/ 32chs | |
| 2µs | 1.1min | 2.1sec | 9.7hrs | - | - | - |
| 5µs | 2.8min | 5.2sec | 24hrs | - | - | - |
| 20µs | 11min | 21sec | 4days | 3hrs | - | - |
| 50µs | 28min | 52sec | 10days | 7.6hrs | - | - |
| 100µs | 56min | 1.8min | 20days | 15hrs | - | - |
| 500µs | 4.7hrs | 8.7min | 101days | 3.2days | - | - |
| 1ms | 9.3hrs | 17min | 203days | 6.3days | 3days | 2hrs |
| 50ms | 19days | 15hrs | 28years | 316days | 148days | 4.6days |
| 100ms | 39days | 29hrs | 56years | 1.7years | 296days | 9.3days |
Above data is calculated value.
- *1: Value at sample filing by integers. At peak filing, the recordable time becomes half of above data.
PC Software

- Unifizer: NS3000 Series
- Enables remote control of DL2800 by PC.
- Real-time data transfer from DL2800 to the PC via Ethernet.
(Max speed: 100µs. Up to 8 DL2800s can be connected.) - Display of replayed data (waveform, X-Y, physical data)
- Recorded data can be converted to CSV and output to report easily.
- Various math functions:
Calculations between channels, power, square root, absolute value, common logarithm, exponential, RMS, trigonometric function,moving average, etc.
Main Unit & Accessories
| Item | Model | Specification |
|---|---|---|
| Logger Station | DL2800 |
Input Amplifiers Unit
| Item | Model | Specification |
|---|---|---|
| 2ch High Resolution DC Amp | AP11-101 | Input: ±100mV to ±500V, A/D: 16 bits |
| 2ch FFT Amp | AP11-102 | Anti-aliasing filter: 72dB/OCT, with power supply for sensor |
| 2ch High speed DC Amp | AP11-103 | Input: ±100mV to ±500V, A/D: 12 bits |
| 2ch AC Strain Amp *1 | AP11-104A | Frequency characteristics: 2kHz, Bridge power: 5kHz |
| Event Amp | AP11-105 | Input: 8 logic (voltage/contact) |
| 2ch TC/DC Amp | AP11-106A | Input: R, T, J, K, W (±100mV to ±50V), A/D resolution: 15 bits |
| TC/DC Amp | AP11-107 | Input: R, T, J, K (±10mV to ±50V), A/D resolution: 14 bits |
| F/V Converter | AP11-108 | Input: 1Hz to 10kHz |
| 2ch Vibration/RMS Amp | AP11-109 | Input: ±100mV to ±500V, w/ power supply for sensor |
| 2ch DC Strain Amp | AP11-110 | Input: 800µε to 20kµε(BV=5V), 2kµε to 50kµε(BV=2V) |
| 2ch Zero Suppression Amp | AP11-111 | Input: ±100mV to ±500V, A/D: 16-bit |
- *1: Optional AC Bridge Power Unit (RA23-116) is required.
Optional Unit
| Item | Model | Specification |
|---|---|---|
| Remote Unit | RA23-112 | Use external signals to start/stop recording and to control other functions |
| RS-232C Unit | RA23-114 | For control by communication command or for shutdown by UPS |
| AC Bridge Power Unit | RA23-116 | AC bridge power unit for 2ch AC Strain Amp (AP11-104A) |
| Synchronization Unit | RA28-111 | For multiple DL2800 synchronization (up to 10 units) |
PC Software:Unifizer
| Item | Model | Specification |
|---|---|---|
| Online Basic Set | NS3100-P01 | For RA1000/2000 series and L2800 Compatible with Windows 2000 & XP |
| Offline Set | NS3200-P01 |
- * Contents of this catalog including specifications are subject to change without prior notice.
-

- To open .pdf documents, you will need the latest Adobe Acrobat Reader software.
Download Adobe software from left.
If you cannot open the document on your web-browser,
right-click the link and choose option 'Save Target As' to download the pdf into your local storage medium, first.
Then, launch the saved file from your storage device (i.e. HDD of your PC etc.).